Project protocol: Rotarod test
Introduction
The rotarod test assesses motor coordination, motor learning, and balance.
Study design
Paradigm: Baseline study
Panel: Collaborative Cross
Number of strains: 18
Sex: Both
Age: 10-12 wks
Sample size: Seven mice per sex from at least two litters
Number of cohorts: 1
Workflow
Step |
Description |
Equipment |
Data collected |
1 |
Mice are acclimated to test room for at least 30 min |
- |
- |
2 |
Each mouse is placed on the rotarod in a separate lane |
Rotarod |
- |
3 |
Rotarod testing (5 trials) |
Rotarod |
Latency to fall from rotarod |
Equipment, software and supplies
Reagents and solutions
Procedure
- Mice are allowed to acclimate to the testing room for a minimum of 30 min.
- Mice are tested for 5 trials with an acceleration of 7.2 rpm/min (trial max = 5 min).
- Mice are placed in separate compartments of the rotating rod.
- The latency to fall into the tray for each mouse is recorded by photobeam break.
- Mice remain in the tray for the remainder of the trial.
- A 30 s intertrial interval begins after the last mouse falls.
- Equipment is cleaned between subjects with 1% Virkon S.
Data collection: Local data processing is implemented to calculate derived measures from this data and a table of the derived measures and raw data are reported per mouse in JaxLIMS.
Data QC: Failure to stay on wheel at start of test, mice holding any damaged edges of the sandpaper, equipment fault, trial interrupted, and calibration/recording errors are possible reasons to exclude data from analysis.
Measurements
LIMS field name |
MPD variable name |
Description |
Units |
Series |
Primary measurement |
Series order |
MPD computed |
Calculation |
Method |
Data type |
Duration 1 |
latency_trial1 |
latency to fall from rotarod, trial 1 |
s |
Yes |
latency_trial1 |
1 |
|
|
rotarod |
Float |
Duration 2 |
latency_trial1 |
latency to fall from rotarod, trial 2 |
s |
Yes |
latency_trial1 |
2 |
|
|
rotarod |
Float |
Duration 3 |
latency_trial1 |
latency to fall from rotarod, trial 3 |
s |
Yes |
latency_trial1 |
3 |
|
|
rotarod |
Float |
Duration 4 |
latency_trial1 |
latency to fall from rotarod, trial 4 |
s |
Yes |
latency_trial1 |
4 |
|
|
rotarod |
Float |
Duration 5 |
latency_trial1 |
latency to fall from rotarod, trial 5 |
s |
Yes |
latency_trial1 |
5 |
|
|
rotarod |
Float |
Experimental and technical parameters
Equipment name |
Rotarod (accelerating) |
Equipment manufacturer |
Med Associates |
Equipment model |
SOF-ENV-575 |
Diameter of rod |
70 mm |
Material of rod cover |
320 grit sandpaper (applied with rubber cement) |
Start speed |
4 rpm |
Acceleration |
7.2 rpm/min |
Max trial duration |
5 min |
Start time |
0700 - 1800 EST |
Disinfectant |
1% Virkon Solution |
|